Evidence · Peer-reviewed study + hospital field validation

UV-C disinfection robot, validated by Leiden University and field-tested in an ICU.

Leiden University’s Institute of Biology tested Omnia® against three indicator bacteria. The results are published in the international journal Infection Prevention in Practice. A technical field validation at Clemenceau Medical Center (CMC) in Dubai shows which UV-C doses Omnia® delivers in a fully equipped intensive care room.

Infection Prevention in Practice 8 (2026) 100582 · Open access (CC BY 4.0)

Usability and microbial validation study of an autonomous UV-C disinfection robot

T.R. de Jager (Ryberg B.V.), T.O. Delzenne, D. Claessen (Institute of Biology Leiden) · Published online 22 August 2026

DOI: 10.1016/j.infpip.2026.100582

Key results

More than 5 to more than 7 log.

A 5-log reduction means 99.999% of bacteria eliminated. With autonomous (moving) disinfection, Omnia® reached more than 5 log for all three test bacteria in the Leiden study. For comparison, the study notes that traditional disinfectant wipes have been tested up to a 4-log reduction.

>7 log
E. coli · Gram-negative model pathogen
>5 log
P. putida · model for P. aeruginosa
>7 log
B. subtilis · Gram-positive model organism

Evidence at a glance

From controlled lab to intensive care.

The Leiden lab study shows what Omnia® achieves against bacteria. The field validation at CMC shows which UV-C doses Omnia® delivers in a fully equipped ICU room. Two different studies, each with its own scope.

Leiden University studyICU field validation · Clemenceau Medical Center, Dubai
SettingFurnished laboratory set up as a hospital room (~53 m²)Fully equipped ICU room with bed, dialysis machine, respirator and monitors
MeasuredBacterial log reduction, UV-C dose, usabilityUV-C dose at 14 high-touch points, surface cultures, cycle time
Carried out byInstitute of Biology Leiden (independent microbiology)Ryberg engineers with CMC’s infection control officer and medical staff
StatusPeer-reviewed, open access (2026)Ryberg technical report TR002, not peer-reviewed (2024)

In practice

Built for the operating theatre.

Ryberg Omnia® autonomous UV-C disinfection robot with its UV-C lamps switched on in an empty hospital operating room, seen from above, next to the operating table, surgical lights and anaesthesia equipment
Omnia® disinfecting an unoccupied operating theatre after routine cleaning. The Leiden study itself was carried out in a furnished laboratory set up to emulate a hospital room.

Summary · Leiden University study

What the Leiden study shows.

Researchers placed bacteria on high-touch locations in a furnished laboratory set up to emulate a hospital room, then let Omnia® disinfect the space. UV-C doses were measured with independent dosimeters and compared with the robot’s own disinfection reports and the actual bacterial kill.

With autonomous moving disinfection, Omnia® reached more than 5 log for all three organisms. At the same duration and light intensity, moving disinfection outperformed the same robot used statically at four fixed positions: fewer shadows, shorter distances and more even coverage. This is a like-for-like comparison; static devices on the market are often more powerful.

The study concludes that, contrary to earlier research, the technology is ready for daily cleaning and disinfection protocols.

Video · Moving vs static

The study in 100 seconds.

Explainer video (1:40) with voice-over and optional captions: moving versus static UV-C in a patient room and an operating theatre, and the results of the Leiden University study. Room layouts and dose maps are illustrative; the study was carried out in a furnished laboratory set up as a hospital room.

Compare static devices, autonomous robots and Omnia® →

Leiden study · The data

Log reductions per sample location.

Location B was chosen for its partial shadow, exactly where a moving robot makes the difference.

Organism Mode Duration A B C
E. coliStatic (4 positions)13 min>4>3>5
E. coliDynamic · Quick13 min>7 ✱>7 ✱>7 ✱
E. coliStatic (4 positions)28 min>8 ✱>7>7
E. coliDynamic · Normal26 min>6>8 ✱>8
P. putidaDynamic · Normal27 min>5>5>5
B. subtilisDynamic · Deep40 min>8>7>7

Microbial log reduction per sample location. ✱ = no surviving colonies detected. Source: de Jager et al. (2026), Table II. Room area ≈ 53 m².

Omnia disinfection report showing a heat map of UV-C dose across the room and expected log reductions per contaminant
Every cycle ends with an automatic disinfection report: a dose heat map of the room plus the expected log reduction per pathogen. The Leiden study found actual results equal to or better than reported.
Bacillus subtilis culture plates: untreated control with full growth, and three plates after Omnia UV-C disinfection with no growth

Leiden study · Before & after

B. subtilis: no growth after one cycle.

Top left: the untreated control plate, full of colonies. The other three plates were placed at locations A, B and C during a single deep Omnia® cycle, with more than 7 log reduction at every location.

Leiden study · Usability

Predictable time, minimal effort.

The Leiden study derived a simple rule of thumb: about 15 seconds per m² per intensity level.

4–7 min

set-up time by a novice operator, including booting robot and tablets.

13 min

Quick (10 mJ/cm²) for ~53 m² · 14.8 s/m².

27 min

Normal (20 mJ/cm²) for ~53 m² · 30.1 s/m².

40 min

Deep (30 mJ/cm²) for ~53 m² · 45.2 s/m².

Nurse monitoring the Ryberg Omnia® UV-C disinfection robot through the window from outside an empty hospital patient room
Staff start and monitor Omnia® from outside the room while the UV-C cycle runs in the empty patient room.

Field validation · Intensive care unit · CMC Dubai

From the lab to the ICU.

After the lab, a technical field validation in a hospital environment: a fully equipped intensive care room at Clemenceau Medical Center (CMC) in Dubai, with a bed, dialysis machine, respirator, monitors, curtains and overhanging equipment. Together with CMC’s infection control officer, Ryberg engineers selected 14 high-touch points, including bed rails, the bed surface, the blood pressure cuff, the dialysis machine, the respirator and the sink tap. At each point, UV-C dosimeters recorded the dose Omnia® delivered. Omnia® mapped the room itself and then ran two consecutive cycles on a setting that targets 25 mJ/cm².

Ryberg Omnia® white conical UV-C disinfection robot with UV-C lamps switched on in an empty, fully equipped intensive care room at Clemenceau Medical Center in Dubai, with bed, monitors and ventilation equipment
Omnia® disinfecting an unoccupied ICU room at Clemenceau Medical Center (CMC), Dubai, UAE.

12 min

average cycle time for the ICU room (11 and 13 minutes), within the 15-minute target.

86%

of measuring points reached the 25 mJ/cm² target after a single cycle (12 of 14). After the second cycle: 13 of 14.

Negative

surface cultures taken after the first cycle, within the hospital’s acceptance level.

Bar chart of average UV-C dose by exposure group after one and two Omnia cycles in an ICU room: direct exposure 81 and 118, partial exposure 48 and 75, partial shadow 34 and 55, full shadow 10 and 15 mJ/cm²
Average UV-C dose per exposure group, 14 measuring points. Source: Ryberg technical report TR002 (2024).

What the ICU adds to the Leiden findings.

  • Line of sight is predictable. The measured dose followed a clear, near-linear pattern: from direct exposure to partial exposure, partial shadow and full shadow. That makes it possible to plan a cycle around a room’s layout. It also matches the Leiden finding that moving disinfection gives more even coverage than the same device used statically.
  • A second pass fits the existing routine. The only point that stayed below target sat behind the curtains. In daily terminal cleaning, staff already move and turn furniture and equipment. Repositioning items slightly before a second cycle reaches those spots without changing the workflow.
  • Reporting stays on the safe side. The robot’s software assumes worst-case horizontal surfaces, and in this room vertical surfaces received more UV-C than that. The Leiden study likewise found that actual bacterial kill equalled or exceeded what the robot’s report predicted.
UV-C dosimeter card before exposure (yellow) and after an Omnia cycle (pink), next to the colour reference chart from 25 to 100 mJ/cm²
Left: unexposed dosimeter (yellow). Right: after Omnia® cycles (pink). Independent UV-C dosimeters change colour with the dose received; calibrated digital dosimeters were used alongside the stickers.

Beyond the wipe benchmark.

CMC uses universal disinfectant wipes in its cleaning and disinfection routine. Their manufacturer reports a 4-log (99.99%) reduction against SARS-CoV-2 and MERS-CoV. Based on the UV-C doses measured across all points in the ICU room, the calculated reduction for these two viruses was 4.5 log and more than 5 log respectively. Omnia® adds that extra layer on top of the existing routine.

The virus reductions for Omnia® are calculated from the measured UV-C dose using published dose-response data. They were not measured with live viruses. Wipe figures: manufacturer’s published efficacy data.

12 minutes on top of the routine.

At CMC, manual terminal cleaning and disinfection of the ICU room takes two trained housekeeping staff one hour. When terminal fogging is added, that takes about 10 minutes to set up, plus roughly two hours of fogging, evaporation and ventilation. An Omnia® cycle added an average of 12 minutes.

CMC staff took part in demonstrations, training sessions and safety briefings as part of the validation, and operated the robot themselves throughout.

About this field validation.

This field validation was a technical validation in a hospital environment, not a clinical study, and it was not designed to show effectiveness in daily practice. It is a technical report by Ryberg engineers (TR002, March 2024), carried out in close collaboration with Clemenceau Medical Center; results were interpreted together with CMC medical staff. It has not been peer-reviewed. It covers a single ICU room with one series of measurements, using a pre-release version of Omnia® (hardware and software version 1.0 beta). The report measured UV-C dose, not bacterial kill; for bacterial reductions, see the Leiden University study above. UV-C complements manual cleaning and does not replace it.

Independence

Designed and executed independently.

The microbial experiments were designed, executed and quantified independently at Leiden University. Leiden University received no compensation, has no financial interest in Ryberg, and did not acquire a robot from Ryberg. Ryberg’s co-author (engineer T.R. de Jager) is disclosed in the paper’s conflict-of-interest statement.

Scope

What was tested, and what wasn’t.

Tests used safe (ML-1) model organisms representing key hospital pathogens, on smooth, non-porous surfaces in a furnished laboratory rather than a patient room. UV-C complements manual cleaning; it does not replace it. Further testing on more pathogens and shadowed areas is planned.

How to cite

de Jager TR, Delzenne TO, Claessen D. Usability and microbial validation study of an autonomous UV-C disinfection robot. Infection Prevention in Practice 2026;8:100582. https://doi.org/10.1016/j.infpip.2026.100582 · PubMed: PMID 42831187 · PubMed Central: PMC13634504

Open access article under the CC BY 4.0 licence, published by Elsevier Ltd on behalf of The Healthcare Infection Society. Figures on this page are reproduced from the article.