Evidence · Peer-reviewed study + hospital field validation
Leiden University’s Institute of Biology tested Omnia® against three indicator bacteria. The results are published in the international journal Infection Prevention in Practice. A technical field validation at Clemenceau Medical Center (CMC) in Dubai shows which UV-C doses Omnia® delivers in a fully equipped intensive care room.
Key results
A 5-log reduction means 99.999% of bacteria eliminated. With autonomous (moving) disinfection, Omnia® reached more than 5 log for all three test bacteria in the Leiden study. For comparison, the study notes that traditional disinfectant wipes have been tested up to a 4-log reduction.
Evidence at a glance
The Leiden lab study shows what Omnia® achieves against bacteria. The field validation at CMC shows which UV-C doses Omnia® delivers in a fully equipped ICU room. Two different studies, each with its own scope.
| Leiden University study | ICU field validation · Clemenceau Medical Center, Dubai | |
|---|---|---|
| Setting | Furnished laboratory set up as a hospital room (~53 m²) | Fully equipped ICU room with bed, dialysis machine, respirator and monitors |
| Measured | Bacterial log reduction, UV-C dose, usability | UV-C dose at 14 high-touch points, surface cultures, cycle time |
| Carried out by | Institute of Biology Leiden (independent microbiology) | Ryberg engineers with CMC’s infection control officer and medical staff |
| Status | Peer-reviewed, open access (2026) | Ryberg technical report TR002, not peer-reviewed (2024) |
In practice
Summary · Leiden University study
Researchers placed bacteria on high-touch locations in a furnished laboratory set up to emulate a hospital room, then let Omnia® disinfect the space. UV-C doses were measured with independent dosimeters and compared with the robot’s own disinfection reports and the actual bacterial kill.
With autonomous moving disinfection, Omnia® reached more than 5 log for all three organisms. At the same duration and light intensity, moving disinfection outperformed the same robot used statically at four fixed positions: fewer shadows, shorter distances and more even coverage. This is a like-for-like comparison; static devices on the market are often more powerful.
The study concludes that, contrary to earlier research, the technology is ready for daily cleaning and disinfection protocols.
Video · Moving vs static
Leiden study · The data
Location B was chosen for its partial shadow, exactly where a moving robot makes the difference.
| Organism | Mode | Duration | A | B | C |
|---|---|---|---|---|---|
| E. coli | Static (4 positions) | 13 min | >4 | >3 | >5 |
| E. coli | Dynamic · Quick | 13 min | >7 ✱ | >7 ✱ | >7 ✱ |
| E. coli | Static (4 positions) | 28 min | >8 ✱ | >7 | >7 |
| E. coli | Dynamic · Normal | 26 min | >6 | >8 ✱ | >8 |
| P. putida | Dynamic · Normal | 27 min | >5 | >5 | >5 |
| B. subtilis | Dynamic · Deep | 40 min | >8 | >7 | >7 |
Microbial log reduction per sample location. ✱ = no surviving colonies detected. Source: de Jager et al. (2026), Table II. Room area ≈ 53 m².
Leiden study · Before & after
Top left: the untreated control plate, full of colonies. The other three plates were placed at locations A, B and C during a single deep Omnia® cycle, with more than 7 log reduction at every location.
Leiden study · Usability
The Leiden study derived a simple rule of thumb: about 15 seconds per m² per intensity level.
set-up time by a novice operator, including booting robot and tablets.
Quick (10 mJ/cm²) for ~53 m² · 14.8 s/m².
Normal (20 mJ/cm²) for ~53 m² · 30.1 s/m².
Deep (30 mJ/cm²) for ~53 m² · 45.2 s/m².
Field validation · Intensive care unit · CMC Dubai
After the lab, a technical field validation in a hospital environment: a fully equipped intensive care room at Clemenceau Medical Center (CMC) in Dubai, with a bed, dialysis machine, respirator, monitors, curtains and overhanging equipment. Together with CMC’s infection control officer, Ryberg engineers selected 14 high-touch points, including bed rails, the bed surface, the blood pressure cuff, the dialysis machine, the respirator and the sink tap. At each point, UV-C dosimeters recorded the dose Omnia® delivered. Omnia® mapped the room itself and then ran two consecutive cycles on a setting that targets 25 mJ/cm².
average cycle time for the ICU room (11 and 13 minutes), within the 15-minute target.
of measuring points reached the 25 mJ/cm² target after a single cycle (12 of 14). After the second cycle: 13 of 14.
surface cultures taken after the first cycle, within the hospital’s acceptance level.
CMC uses universal disinfectant wipes in its cleaning and disinfection routine. Their manufacturer reports a 4-log (99.99%) reduction against SARS-CoV-2 and MERS-CoV. Based on the UV-C doses measured across all points in the ICU room, the calculated reduction for these two viruses was 4.5 log and more than 5 log respectively. Omnia® adds that extra layer on top of the existing routine.
The virus reductions for Omnia® are calculated from the measured UV-C dose using published dose-response data. They were not measured with live viruses. Wipe figures: manufacturer’s published efficacy data.
At CMC, manual terminal cleaning and disinfection of the ICU room takes two trained housekeeping staff one hour. When terminal fogging is added, that takes about 10 minutes to set up, plus roughly two hours of fogging, evaporation and ventilation. An Omnia® cycle added an average of 12 minutes.
CMC staff took part in demonstrations, training sessions and safety briefings as part of the validation, and operated the robot themselves throughout.
About this field validation.
This field validation was a technical validation in a hospital environment, not a clinical study, and it was not designed to show effectiveness in daily practice. It is a technical report by Ryberg engineers (TR002, March 2024), carried out in close collaboration with Clemenceau Medical Center; results were interpreted together with CMC medical staff. It has not been peer-reviewed. It covers a single ICU room with one series of measurements, using a pre-release version of Omnia® (hardware and software version 1.0 beta). The report measured UV-C dose, not bacterial kill; for bacterial reductions, see the Leiden University study above. UV-C complements manual cleaning and does not replace it.
Independence
The microbial experiments were designed, executed and quantified independently at Leiden University. Leiden University received no compensation, has no financial interest in Ryberg, and did not acquire a robot from Ryberg. Ryberg’s co-author (engineer T.R. de Jager) is disclosed in the paper’s conflict-of-interest statement.
Scope
Tests used safe (ML-1) model organisms representing key hospital pathogens, on smooth, non-porous surfaces in a furnished laboratory rather than a patient room. UV-C complements manual cleaning; it does not replace it. Further testing on more pathogens and shadowed areas is planned.
How to cite
Open access article under the CC BY 4.0 licence, published by Elsevier Ltd on behalf of The Healthcare Infection Society. Figures on this page are reproduced from the article.
In the media
Robotics & Automation News · 28 October 2026 · English
Sam Francis · Read the article →
Health&Care Management (HCM) · 29 October 2026 · German
Redaktion HCM · Read the article →
Schoonmaakjournaal · 28 October 2026 · Dutch
Esmar van Langen · Read the article →
ICT&health · 23 October 2026 · Dutch
Redactie ICT&health · Read the article →
The full peer-reviewed paper as a PDF, including methods, figures and references.
Validated by Leiden University and field-tested in an intensive care unit. Request an on-site demo and we will map one of your rooms together.